ISR 6 Advanced Manual Data format UPP
Ò
· 40
Description Command Parameters
Measuring value
(one-channel and
ratio temperature)
AAek Answer: SSSSSQQQQQ
2x5 decimal digits (in °C or °F, last digit is 1/10 °C or °F),
SSSSS = one-channel temperature
QQQQQ=ratio temperature
Emissivity e for one-
channel temperature
AAemXXXX
AAem
XXXX=0050 to 1000 e=0.050 to 1.000
Answer: DDDD 4 decimal digits 0050 to 1000
Transmittance t of
window
AAetXXXX
AAet
XXXX=0050 to 1000 t=0.050 to 1.000
Answer: DDDD 4 decimal digits 0050 to 1000
K = e
1
/ e
2
Emissivity ratio
AAevXXXX
AAev
XXXX=0800 to1200 e
1
/ e
2
=0.800 to 1.200
Answer: DDDD 4 decimal digits 0800 to 1200
Response time t
90
AAezX X=0 to 6 0=min. 1=0.01 s 2=0.05 s
4=1.00 s 5=3.00 s 6=10
Temp. Display °C or °F AAfhX Output: X = 0 display in °C
X = 1 display in °F
Device Address: AAgaXX XX = (00 to 97)
00 to 97 = regular device addresses
98 = global address with response
99 = global address with response
internal temperature
(read)
AAgt
AAtm
Answer: DDD 3 decimal digits (000 to 098 °C or
032 to 210 °F)
gt=current temp. tm=maximum temp. (memory)
Operation mode AAkaX X = 1 mono mode (1 channel)
X = 2 ratio mode (2 channel)
Laser AAlaX
AAla
X=0 off X=1 on
Answer: 1 digit: “0” or “1”
Software simulation
of external clearance
AAlx Clears maximum storage
Clear peak memory
t
clear
AAlzX X=0 to 8 0= OFF 1=0.01s 2=0.05s 3=0.25s
4=1.0s 5=5.0s 6=25.0s
8=AUTO 9=HOLD
Basic range (read) AAmb Answer: XXXXYYYY
2x4 hex-digit for lower and upper range limit (°C or °F)
Sub range (read) AAme Answer: XXXXYYYY
2x4 hex-digit for lower and upper range limit (°C or °F)
Sub range (set) 1. AAm1XXXXYYYY XXXXYYYY=2x4 hex-digit for lower and upper sub range
limit (°C or °F)
2. AAm2 AAm2 confirms the change (auto reset)
Measuring value AAms Answer: QQQQQ (88880=Overflow)
5 decimal digit (in °C or °F, last digit is 1/10 °C or °F)
Device type AAna Output: “ISR 6 Advanced ” (16 ASCII-characters)